Human anti-HIV IgM detection by the OraQuick ADVANCE® Rapid HIV 1/2 Antibody Test

View article
PeerJ
A correction to this article was published on June 22, 2018. Read the correction.

Introduction

The Centers for Disease Control and Prevention (CDC) and many public health jurisdictions continue to advocate for the most sensitive rapid HIV test that is available, both to identify positive patients when they are at their most infectious, and to ensure that negative patients are truly negative when starting a pre-exposure prophylaxis regimen (Centers for Disease Control and Prevention and Association of Public Health Laboratories. Laboratory Testing for the Diagnosis of HIV Infection: Updated Recommendations, 2014). Currently, the recommendation is to utilize tests that can detect HIV infection biomarkers within 30 days of infection, when initial immune responses are mounted. Typically, these biomarkers measure the presence of virus, such as p24 antigen or viral RNA, the patient’s IgM and IgG immune response to HIV infection, and/or the level of CD4 T cells (see Fig. 1).

Various biomarkers utilized for HIV detection.

Figure 1: Various biomarkers utilized for HIV detection.

Stages of HIV life cycle (Fiebig stages) and relative abundance of biomarkers throughout these stages. Different biomarkers allow for detection of infection at different stages, represented by the boxes and labels below the chart. This figure is adapted from Fiebig et al. (2003).

The CDC and the Association of Public Health Laboratories (APHL) have issued revised recommendations (Centers for Disease Control and Prevention and Association of Public Health Laboratories. Laboratory Testing for the Diagnosis of HIV Infection: Updated Recommendations, 2014) based on HIV tests approved by the Food and Drug Administration (FDA). The CDC and APHL recommend that laboratories use an automated, laboratory-based p24 antigen/antibody HIV screening immunoassay followed, if reactive, by an HIV-1/HIV-2 antibody differentiation immunoassay. When the differentiation assay interpretation is negative or indeterminate for HIV-1, an HIV-1 nucleic acid test (NAT) is recommended (see Fig. 2).

Centers for disease control and prevention HIV testing algorithm.

Figure 2: Centers for disease control and prevention HIV testing algorithm.

Reproduced from laboratory testing for the diagnosis of HIV infection: updated recommendations, 2014.

For those settings where an automated laboratory instrument is not available or serum/plasma sample collection is not practical or cost-effective to utilize, a separate guideline for testing in non-clinical settings has been established (Centers for Disease Control and Prevention. Implementing HIV Testing in Nonclinical Settings: A Guide for HIV Testing Providers, 2016). These guidelines recommend that sites use HIV testing technologies that are the most sensitive, cost-effective, and feasible for use by their agency.

Rapid, POC testing is a logical choice for screening in a non-clinical setting. It is important, for the reasons outlined above, that a POC test is still able to detect infection within the 30-day acute infection window period. The only way that this can occur is by the detection of either the IgM antibody response or p24 antigen (Fiebig et al., 2003). While there is an FDA-approved POC test that detects p24 antigen available today, numerous data has been presented and published showing limited utility of p24 detection in fingerstick whole blood samples due to low sensitivity of the device in the 20–50 pg/ml range compared to automated p24/IgM/IgG assays with sensitivity as low as 0.1 pg/ml (Pebody, 2012; Conway et al., 2014; Lewis et al., 2015). This lack of sensitivity drives the market towards POC tests that detect IgM antibodies.

A recent study has demonstrated that POC tests using a Protein A-based detection system can detect samples with predominantly HIV-1 IgM reactivity (Moshgabadi et al., 2015). The OraQuick ADVANCE® Test also uses Protein A as the detection protein in the antibody-binding colloidal gold conjugate. Due to this, it is expected that the OraQuick ADVANCE® Test will also detect samples with predominantly IgM reactivity.

The OraQuick ADVANCE® Test has proven performance and patient acceptability that is facilitated by specimen flexibility (Wesolowski et al., 2006, Pant Pai et al., 2007, 2012; Zachary et al., 2012, Reynolds & Muwonga, 2014, Nkenfou et al., 2014; Martin et al., 2017). The OraQuick ADVANCE® Test is a simple, three-step protocol providing results in 20 minutes. The test is FDA-approved with >99% sensitivity and specificity claimed across many specimen types, including oral fluid, fingerstick blood and whole blood. FDA approval for the OraQuick ADVANCE® Test using oral fluid as the specimen type was in 2004, and raw material and manufacturing process changes that allowed earlier detection of seroconversion was FDA approved in 2015.

This report describes the results of a study designed to evaluate the seroconversion performance of the OraQuick ADVANCE® Test compared to automated IgG/IgM enzyme immunoassay (EIA) performance and to confirm the hypothesis that its Protein A-based detection system is able to detect samples known to be predominantly IgM positive, thereby demonstrating it is appropriate as a POC diagnostic test for acute HIV infection.

Materials and Methods

OraQuick ADVANCE® Rapid HIV-1/2 Antibody Test

The OraQuick ADVANCE® Test is FDA approved to detect antibodies to human immunodeficiency virus type 1 (HIV-1) and type 2 (HIV-2) in whole blood (from a fingerstick or venipuncture sample), plasma, or oral fluid sample. The assay strip in the device consists of a membrane of nitrocellulose striped with control and test lines. The test line is striped with both HIV-1 and HIV-2 synthetic peptides and modified streptavidin (a biotin-binding protein). HIV-specific antibodies are introduced with the sample, and, if present, will interact with biotinylated HIV-1 and HIV-2 peptides and colloidal gold labeled with Protein A, which are impregnated in different sections of the device. This sample-reagent mixture then flows up the nitrocellulose, with complexed gold particles binding to the test line components in various “sandwich” configurations (Fig. 3).

Operating principle of the OraQuick ADVANCE® Test.

Figure 3: Operating principle of the OraQuick ADVANCE® Test.

Sample is wicked up the device, hydrating blocker and conjugate pad reagents, which are transferred to the nitrocellulose membrane. HIV antibodies in the sample bind biotinylated reagents and are labeled by Protein A colloidal gold. The immuno-sandwich is immobilized by biotin binding with Streptavidin-BSA, and by anti-HIV antibodies interacting with anchored peptides at the test line. Human antibodies complexed with Protein A colloidal gold are captured at the control line. Accumulation of colloidal gold at the control and test lines creates a reddish-purple stripe.

The presence of HIV-1 and/or HIV-2 antibodies in the tested specimen is indicated by the appearance of a reddish-purple colored band at the test line location of the device within 20–40 minutes.

Comparison of seroconversion detection by OraQuick ADVANCE® Rapid HIV-1/2 Antibody Test to IgG/IgM automated assays

The performance of three independent production lots of the OraQuick ADVANCE® Test (lot numbers 15070PL0, 6650009, and 6650592) was evaluated across 201 samples from 23 different commercially available seroconversion panels (Table 1), as described in the package insert. Each panel consists of sequential plasma specimens obtained from a single HIV-infected individual during seroconversion. Five microliters of each seroconversion panel member was spiked into developer solution and three replicates were run per panel member for each lot. Results were read after a 20 minute incubation period. The results were compared to those provided by the seroconversion panel vendor, which were determined using FDA-approved automated IgG/IgM assays.

Table 1:
Commercially available seroconversion panels used in analysis of seroconversion sensitivity for the OraQuick ADVANCE® test.
Vendor Panel name Panel number(s)
SeraCare PRB924 1–8
SeraCare PRB925 1–6
SeraCare PRB926 1–6
SeraCare PRB933 1–3
SeraCare PRB934 1–3
SeraCare PRB939 1–9
SeraCare PRB945 1–6
SeraCare PRB947 1–4
SeraCare PRB959 1–7
SeraCare PRB965 1–6
SeraCare PRB967 1–6
SeraCare PRB968 1–10
SeraCare PRB969 1–10
SeraCare PRB970 1–4
ZeptoMetrix 6243 1–10
ZeptoMetrix 9015 1–10
ZeptoMetrix 9019 1–3
ZeptoMetrix 9032 1–14
ZeptoMetrix 9076 1–10
ZeptoMetrix 9077 1–29
ZeptoMetrix 9079 1–25
ZeptoMetrix 9084 1–4
ZeptoMetrix 12008 1–13
DOI: 10.7717/peerj.4430/table-1

Detection of IgM antibodies

A previous study characterized samples in commercially available seroconversion panels as containing detectable levels of either IgG-only, IgG and IgM, or IgM-only by depleting all IgM antibodies from the sample then testing with a reference ELISA and a POC antibody test (Moshgabadi et al., 2015). Due to limited sample availability from the vendor, 15 of the 20 samples identified as being predominantly IgM (greater than 50% of total immunoglobulin content) were utilized in this study. These samples were tested concurrently with the OraQuick ADVANCE® Test, and the INSTI® HIV-1/HIV-2 antibody test (INSTI™; bioLytical Laboratories, Richmond, BC, USA) as described in the original report (Moshgabadi et al., 2015), following the package insert in each case.

In this study, these samples were evaluated using the OraQuick ADVANCE® Test by spiking 5 microliters of each sample into developer solution. The device was read after a 20 minute incubation period as per the package insert. The INSTI® assay required 50 microliters of sample and the protocol was followed as per the package insert.

Results

Comparison to IgG/IgM automated assays

A total of 201 panel members were tested across the 23 seroconversion panels. Each panel member was run in triplicate and a result was determined to be positive with the presence of a complete test line at or above the limit of visual detection in all three replicates (see Table 2). The average performance of the OraQuick ADVANCE® Test was compared to the automated EIA data reported in the seroconversion package inserts. Thirteen of the 23 panels had data reported from more than one EIA platform, and in these cases the average of the two or three results was used to compare to the OraQuick ADVANCE® Test results. A total of 114 of the 201 panel members tested negative for both EIA and in all three lots of the OraQuick ADVANCE® Test. When comparing the overall average of seroconversion detection days after first bleed for EIA to the three lot average, the OraQuick ADVANCE® Test detected seroconversion within 1.5 days of the automated IgG/IgM assays (95% CI [−0.1 to 3.1 days]), with three panels being detected earlier by the OraQuick ADVANCE® Test, seven on the same day and 13 later.

Table 2:
Comparison of HIV seroconversion detection between three lots of the OraQuick ADVANCE® HIV Test and automated enzyme immunoassay (EIA) tests.
Panel Days after initial bleed Days OraQuick ADVANCE average–EIA average
OraQuick 15070PL0 OraQuick 6650009 OraQuick 6650592 OraQuick average Abbott HIV 1/2 EIA Abbott HIV AB 1/2 AxSYM EIA Abbott prism anti HIV 1/2 EIA Genetic systems HIV-1/2 EIA Bio-rad HIV-1/2 plus O EIA EIA average
924 35 40 35 37 33 33 3.7
925 44 44 44 44 44 49 46.5 −2.5
926 27 27 27 27 27 27 27 0.0
933 27 27 27 27 21 27 24 3.0
934 7 7 7 7 7 11 9 −2.0
939 103 103 103 103 103 103 103 0.0
945 20 20 20 20 13 13 7.0
947 11 20 20 17 9 20 14.5 2.5
959 14 14 14 14 9 14 11.5 2.5
965 12 12 12 12 12 12 0.0
967 19 19 19 19 17 17 2.0
968 33 33 33 33 26 26 7.0
969 70 70 70 70 70 70 0.0
970 14 14 14 14 10 10 4.0
6243 32 32 32 32 32 32 32 0.0
9015 35 35 35 35 35 37 30 34 1.0
9019 38 38 38 38 38 38 38 38 0.0
9032 36 36 36 36 24 51 29 34.7 1.3
9076 74 69 69 71 66 66 66 4.7
9077 45 45 45 45 57 52 54.5 −9.5
9079 55 55 55 55 47 47 47 8.0
9084 49 49 49 49 49 49 0.0
12008 33 35 35 34 33 33 33 1.3
Average 36.2 36.7 36.5 36.5 34.8 32.0 32.0 37.7 38.0 35 1.5
OQ average for equivalent samples 37.3 32.0 32.0 34.8 40.9
OQ average–EIA average for equivalent samples 2.4 0.0 0.0 −2.9 2.9
DOI: 10.7717/peerj.4430/table-2

Detection of IgM antibodies

Among the 15 samples that were characterized as either IgM-only or predominantly IgM (Table 1), 14 of 15 were detectable by both the OraQuick ADVANCE® test and INSTI™ (Table 3). As these samples were shown previously to be reactive (Moshgabadi et al., 2015), any sample that was initially non-reactive was repeated in triplicate to confirm the result. The one sample not detected by the OraQuick ADVANCE® Test was 943-06 (IgM-only), and the one sample not detected by INSTI® was 938-03 (IgM-only). One sample (950-04—predominantly IgM) was initially scored as non-reactive with the OraQuick ADVANCE® Test, but was positive in all three confirmatory tests.

Table 3:
Detection of predominantly IgM or IgM-only samples by the OraQuick ADVANCE® Test and INSTI® test.
Panel name sample Panel number % IgM in sample (%) OraQuick ADVANCE® Rapid HIV-1/2 Antibody Test INSTI® HIV-1/2 antibody test
Result Result
PRB914 1 100 R R
PRB914 2 >50 R R
PRB914 3 100 R R
PRB924 8 >50 R R
PRB925 5 100 R R
PRB925 6 >50 R R
PRB927 3 >50 R R
PRB928 2 100 R R
PRB934 2 >50 R R
PRB934 3 >50 R R
PRB938 3 100 R NR
PRB940 4 >50 R R
PRB943 6 100 NR R
PRB944 5 100 R R
PRB950 4 >50 R R
DOI: 10.7717/peerj.4430/table-3

Note:

Seroconversion panel members from SeraCare were previously determined to have predominantly IgM or IgM-only HIV antibodies (Moshgabadi et al., 2015). Samples were tested with OraQuick ADVANCE® and INSTI® tests and reactivity recorded.

Discussion

The revised HIV testing recommendations published by the CDC and APHL in June, 2014 have focused testing on early detection during the acute phase of infection (Centers for Disease Control and Prevention and Association of Public Health Laboratories. Laboratory Testing for the Diagnosis of HIV Infection: Updated Recommendations, 2014). Lab-based tests typically use automated p24 antigen/antibody HIV screening immunoassays followed, if reactive, by an HIV-1/HIV-2 antibody differentiation immunoassay. The ability to detect p24 antigen and/or IgM allows these assays to detect the acute phase of infection when patients are most infectious.

It is well established that persons at high risk for HIV infection, such as gay and bisexual males, injection drug users, and minority populations have a history of not engaging in routine care for a variety of system, social and individual factors (Reimen et al., 2015). For these populations, as well as those in rural settings, rapid POC HIV diagnostics have been shown to help identify undiagnosed infections and increase screening in all levels of the healthcare system and outreach programs (Tucker, Bien & Peeling, 2013). It is important that these more convenient tests are also able to detect acute infections, either by p24 antigen or IgM antibody detection.

The performance of the OraQuick ADVANCE® Test has been established in clinical population settings to support CE approval in Europe, FDA approval in the US and PreQualification by the World Health Organization. The purpose of the study reported here was to evaluate the ability of the OraQuick ADVANCE® Test to (i) detect seroconversion of plasma panels collected from HIV-infected subjects compared to automated lab-based IgM/IgG assays, and (ii) detect HIV antibodies in plasma samples that contain predominantly IgM antibodies to HIV. It is noted that the authors are not aware of any attempts to replicate these results by others.

The results shown in Table 2 demonstrate that the OraQuick ADVANCE® Test is able to detect seroconversion an average of 1.5 days later than automated assays. While there were differences between platforms, with average differences ranging from 2.9 days earlier (Genetic Systems, Redmond, WA, USA) to 2.9 days later (BioRad, Hercules, CA, USA), 10 of the 23 panels detected seroconversion with the OraQuick ADVANCE® Test at the same time or earlier than the automated assays.

It is possible that these panels, and others, would be detected sooner with a test that also includes p24 antigen detection, which highlights a known limitation of antibody-only testing platforms. However, as described above, the only FDA-approved rapid test with p24 antigen does not have sufficient sensitivity in fingerstick whole blood applications (Smallwood et al., 2016), which is preferred for blood-based testing in non-clinical settings, and drives the POC market towards tests that detect IgM antibodies.

Testing of 15 seroconversion samples that were previously characterized (Moshgabadi et al., 2015) as containing predominantly IgM antibodies with the OraQuick ADVANCE® Test and INSTI® demonstrated both tests are able to detect IgM antibodies, with each device detecting 14 of 15 predominantly IgM samples, although each failed to detect different samples (Table 3).

Conclusion

The results of this study demonstrate that the OraQuick ADVANCE® Test can detect IgM antibodies during an acute infection window period. Based on this capability it can be inferred that the test is able to detect seroconversion approximately 20–25 days after infection, and is therefore suitable for use in testing environments requiring adherence to the CDC and APHL recommendations.

11 Citations   Views   Downloads